PM1605 - Flush pressure sensor
PRODUCENT CODE: PM1605
CATEGORY: Flush transmitters for the process technology
Producer: ifm electronic
2082.09 PLN
(net)
7-14 days
The product is available on request
At the moment, the product is not available in our warehouse.
After placing the order, we will notify you of the expected delivery date of the product.
At the moment, the product is not available in our warehouse.
After placing the order, we will notify you of the expected delivery date of the product.
Fast shipping
Free delivery
from 1000 PLN (net)
Free delivery
from 1000 PLN (net)
Do you have questions about the product?
Ask an experienced consultant:
Ask an experienced consultant:
Description
PM-004-REA01-E-ZVG/US
- Robust stainless steel housing for use in the food and beverage industry
- Hygienic, flush design
- Resistant to high-pressure cleaning with aggressive cleaning agents
- High temperature resistance and high protection rating
IO-Link features
- Precise analogue output and user-friendly communication via IO-Link
Technical data
Product characteristics | |||||||||
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 | ||||||||
Measuring range |
|
||||||||
Process connection | threaded connection G 1 external thread sealing cone | ||||||||
Application | |||||||||
Special feature | Gold-plated contacts | ||||||||
Measuring element | ceramic-capacitive pressure measuring cell | ||||||||
Temperature monitoring | no | ||||||||
Application | flush mountable for the food and beverage industry | ||||||||
Media | viscous media and liquids with suspended particles; liquids and gases | ||||||||
Medium temperature [°C] | -25...150 | ||||||||
Min. bursting pressure |
|
||||||||
Pressure rating |
|
||||||||
Vacuum resistance [mbar] | -1000 | ||||||||
Type of pressure | relative pressure; vacuum | ||||||||
No dead space | yes | ||||||||
MAWP (for applications according to CRN) [bar] | 30 | ||||||||
Electrical data | |||||||||
Operating voltage [V] | 18...30 DC | ||||||||
Min. insulation resistance [MΩ] | 100; (500 V DC) | ||||||||
Protection class | III | ||||||||
Reverse polarity protection | yes | ||||||||
Integrated watchdog | yes | ||||||||
Current consumption [mA] | 3.5...21.5 | ||||||||
Power-on delay time [s] | 1 | ||||||||
Current consumption [mA] | < 45 | ||||||||
Power-on delay time [s] | 0.5 | ||||||||
Inputs / outputs | |||||||||
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 | ||||||||
Outputs | |||||||||
Total number of outputs | 2 | ||||||||
Output signal | analogue signal; IO-Link; (configurable) | ||||||||
Number of digital outputs | 1; (IO-Link) | ||||||||
Number of analogue outputs | 1 | ||||||||
Analogue current output [mA] | 4...20; (scalable) | ||||||||
Max. load [Ω] | 700; (Ub = 24 V; (Ub - 9 V) / 21.5 mA) | ||||||||
Short-circuit proof | yes | ||||||||
Overload protection | yes | ||||||||
Measuring/setting range | |||||||||
Measuring range |
|
||||||||
Analogue start point |
|
||||||||
Analogue end point |
|
||||||||
In steps of |
|
||||||||
Factory setting |
|
||||||||
Accuracy / deviations | |||||||||
Repeatability [% of the span] | < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1) | ||||||||
Characteristics deviation [% of the span] | < ± 0,2; (linearity incl. hysteresis and repeatability, limit value setting to DIN EN IEC 62828-1) | ||||||||
Linearity deviation [% of the span] | < ± 0,15; (Turn down 1:1) | ||||||||
Hysteresis deviation [% of the span] | < ± 0,15; (Turn down 1:1) | ||||||||
Long-term stability [% of the span] | < ± 0,1; (Turn down 1:1; per year) | ||||||||
Total deviation over temperature range |
|
||||||||
Notes on the accuracy / deviation | for further details see section Diagrams and graphs | ||||||||
Response times | |||||||||
Damping for the analogue output dAA [s] | 0...4 | ||||||||
Step response time analogue output [ms] | 30 | ||||||||
Step response time analogue output [ms] | 7 | ||||||||
Interfaces | |||||||||
Communication interface | IO-Link | ||||||||
Transmission type | COM2 (38,4 kBaud) | ||||||||
IO-Link revision | 1.1 | ||||||||
SDCI standard | IEC 61131-9 | ||||||||
Profiles | Digital Measuring Sensor (0x000A), Identification and Diagnosis (0x4000) | ||||||||
SIO mode | no | ||||||||
Required master port type | A | ||||||||
Process data analogue | 3 | ||||||||
Min. process cycle time [ms] | 3.2 | ||||||||
IO-Link resolution pressure [bar] | 0.001 | ||||||||
IO-Link process data (cyclical) |
|
||||||||
IO-Link functions (acyclical) | application specific tag; internal temperature | ||||||||
Supported DeviceIDs |
|
||||||||
Operating conditions | |||||||||
Ambient temperature [°C] | -25...80 | ||||||||
Storage temperature [°C] | -40...100 | ||||||||
Protection | IP 67; IP 68; IP 69K | ||||||||
Tests / approvals | |||||||||
EMC |
|
||||||||
Shock resistance |
|
||||||||
Vibration resistance |
|
||||||||
MTTF [years] | 323 | ||||||||
Note on approval |
|
||||||||
UL approval |
|
||||||||
Mechanical data | |||||||||
Weight [g] | 337.55 | ||||||||
Materials | stainless steel (1.4404 / 316L); PBT | ||||||||
Materials (wetted parts) | ceramics (99.9 % Al2O3); stainless steel (1.4435 / 316L); surface characteristics: Ra < 0,4 / Rz 4; PTFE | ||||||||
Min. pressure cycles | 100 million | ||||||||
Tightening torque [Nm] | 20 | ||||||||
Process connection | threaded connection G 1 external thread sealing cone | ||||||||
Remarks | |||||||||
Pack quantity | 1 pcs. | ||||||||
Electrical connection | |||||||||
Connection | Connector: 1 x M12; coding: A; Contacts: gold-plated | ||||||||
ambient temperature influence on the accuracy |
|