Description
PM-,40BREA01-E-ZVG/US
- Robust stainless steel housing for use in the food and beverage industry
- Hygienic, flush design
- Resistant to high-pressure cleaning with aggressive cleaning agents
- High temperature resistance and high protection rating
IO-Link features
- Precise analogue output and user-friendly communication via IO-Link
Technical data
Product characteristics | |||||||||
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 | ||||||||
Measuring range |
|
||||||||
Process connection | threaded connection G 1 external thread Aseptoflex Vario | ||||||||
Application | |||||||||
Special feature | Gold-plated contacts | ||||||||
Measuring element | ceramic-capacitive pressure measuring cell | ||||||||
Temperature monitoring | no | ||||||||
Application | flush mountable for the food and beverage industry | ||||||||
Media | viscous media and liquids with suspended particles; liquids and gases | ||||||||
Medium temperature [°C] | -25...150 | ||||||||
Min. bursting pressure |
|
||||||||
Pressure rating |
|
||||||||
Vacuum resistance [mbar] | -1000 | ||||||||
Type of pressure | relative pressure | ||||||||
No dead space | yes | ||||||||
Electrical data | |||||||||
Operating voltage [V] | 18...30 DC | ||||||||
Min. insulation resistance [MΩ] | 100; (500 V DC) | ||||||||
Protection class | III | ||||||||
Reverse polarity protection | yes | ||||||||
Measuring principle | hydrostatic | ||||||||
Integrated watchdog | yes | ||||||||
Current consumption [mA] | 3.5...21.5 | ||||||||
Power-on delay time [s] | 1 | ||||||||
Current consumption [mA] | < 45 | ||||||||
Power-on delay time [s] | 0.5 | ||||||||
Inputs / outputs | |||||||||
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 | ||||||||
Outputs | |||||||||
Total number of outputs | 2 | ||||||||
Output signal | analogue signal; IO-Link; (configurable) | ||||||||
Number of digital outputs | 1; (IO-Link) | ||||||||
Number of analogue outputs | 1 | ||||||||
Analogue current output [mA] | 4...20; (scalable) | ||||||||
Max. load [Ω] | 700; (Ub = 24 V; (Ub - 9 V) / 21.5 mA) | ||||||||
Short-circuit proof | yes | ||||||||
Overload protection | yes | ||||||||
Measuring/setting range | |||||||||
Measuring range |
|
||||||||
Analogue start point |
|
||||||||
Analogue end point |
|
||||||||
In steps of |
|
||||||||
Factory setting |
|
||||||||
Accuracy / deviations | |||||||||
Repeatability [% of the span] | < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1) | ||||||||
Characteristics deviation [% of the span] | < ± 0,2 (DIN EN 61298-2); (incl. zero point and span error, non-linearity, hysteresis; Turn down 1:1) | ||||||||
Linearity deviation [% of the span] | < ± 0,15; (Turn down 1:1) | ||||||||
Hysteresis deviation [% of the span] | < ± 0,15; (Turn down 1:1) | ||||||||
Long-term stability [% of the span] | < ± 0,1; (Turn down 1:1; per year) | ||||||||
Total deviation over temperature range |
|
||||||||
Notes on the accuracy / deviation | for further details see section Diagrams and graphs | ||||||||
Response times | |||||||||
Damping for the analogue output dAA [s] | 0...4 | ||||||||
Step response time analogue output [ms] | 30 | ||||||||
Step response time analogue output [ms] | 7 | ||||||||
Interfaces | |||||||||
Communication interface | IO-Link | ||||||||
Transmission type | COM2 (38,4 kBaud) | ||||||||
IO-Link revision | 1.1 | ||||||||
SDCI standard | IEC 61131-9 | ||||||||
Profiles | Digital Measuring Sensor (0x000A), Identification and Diagnosis (0x4000) | ||||||||
SIO mode | no | ||||||||
Required master port type | A | ||||||||
Process data analogue | 3 | ||||||||
Min. process cycle time [ms] | 3.2 | ||||||||
IO-Link resolution pressure [mbar] | 0.1 | ||||||||
IO-Link functions (acyclical) | application specific tag; internal temperature | ||||||||
Supported DeviceIDs |
|
||||||||
Operating conditions | |||||||||
Ambient temperature [°C] | -25...80 | ||||||||
Storage temperature [°C] | -40...100 | ||||||||
Protection | IP 67; IP 68; IP 69K | ||||||||
Tests / approvals | |||||||||
EMC |
|
||||||||
Shock resistance |
|
||||||||
Vibration resistance |
|
||||||||
MTTF [years] | 364 | ||||||||
Mechanical data | |||||||||
Weight [g] | 272.05 | ||||||||
Materials | stainless steel (1.4404 / 316L); PBT | ||||||||
Materials (wetted parts) | ceramics (99.9 % Al2O3); stainless steel (1.4435 / 316L); surface characteristics: Ra < 0,4 / Rz 4; PTFE | ||||||||
Min. pressure cycles | 100 million | ||||||||
Tightening torque [Nm] | 35 | ||||||||
Process connection | threaded connection G 1 external thread Aseptoflex Vario | ||||||||
Displays / operating elements | |||||||||
Display unit | mbar; psi; kPa; mmWS; inH2O | ||||||||
Remarks | |||||||||
Pack quantity | 1 pcs. | ||||||||
Electrical connection | |||||||||
Connection | Connector: 1 x M12; coding: A; Contacts: gold-plated | ||||||||
ambient temperature influence on the accuracy |
|