red-y smart pressure controller
At the moment, the product is not available in our warehouse.
After placing the order, we will notify you of the expected delivery date of the product.
Free delivery
from 1000 PLN (net)
Ask an experienced consultant:
Digital Pressure Controllers for Gases with integrated Flow Measurement
The electronic red-y smart pressure controller combine the reliable technology of our digital mass flow controllers with electronic pressure control. The devices automatically control a predefined process pressure and at the same time measure and/or limit the flow rate. On-the-fly switching between pressure control and flow control offers maximum flexibility.
- 1 Device – 3 Functions
- The pressure controller combines three functions:
- Pressure controller
- Pressure controller with flow measurement/limitation
- Flow controller with pressure measurement
- The following instrument versions are available:
- Integrated pressure control
Accuracy:± 0.5 % of full scale - Integrated back pressure control
Accuracy:± 0.5 % of full scale - Pressure control with external pressure transmitters
- Pressure controller with gas mixer function
- Integrated pressure control
- The pressure controller combines three functions:
-
Analog & Digital: 2 in 1
The pressure controllers make use of the latest CMOS technology and have a digital (Modbus RTU) and analog interface as standard.Profibus, Profinet RT and EtherCAT interfaces are available.
A wide range of cables and power supply units for fast integration of the red-y smart pressure controllers is available - Software get red-y
- Efficient device management with the free get red-y software:
- View flow rate & temperature
- Change set points
- Select measured gas
- Visualization of measured data
- Adjusting control parameter
- Optional modules get red-y software:
- Datalogging
- Gasmixing
- See all available cables and power supply accessories for easy PC connection.
- Efficient device management with the free get red-y software:
- Options
- Built-in display
Display includes: Flow rate, totalizer, unit of measure & set point control - Multigas
One controller or controller can be used for up to 10 different gases or gas mixtures - Profibus
The instruments are available with Profibus interface: DP-V0 & DP-V1 protocols - Industrial Ethernet
Two industrial ethernet protocols Profinet RT and EtherCAT are available - IP67/ATEX Versions
red-y industrial pressure controllers are available as IP67/ATEX versions
- Built-in display
- It‘s a red-y smart
The pressure controllers combine the innovative equipment design of the red-y smart series with the development competence of Vögtlin Instruments GmbH. High-quality components ensure long and trouble-free operation. - Operating Status Indication
The instruments offer a built-in LED status indication - 3-Year Warranty
High-quality components ensure long and trouble-free operation (does not apply to calibration, options and accessories)
Versions
-
red-y smart pressure controller GSP
Digital Pressure Controller with integrated Flow Measurement
Flow up to 60 ln/min
Aluminium version
Body size ¼” -
red-y smart controller GSB
Digital Back Pressure Controller with integrated Flow Measurement
Flow up to 60 ln/min
Aluminium version
Body size ¼” -
Flow up to 450 ln/min
Pressure controller and back pressure controller with measuring ranges up to 450 ln/min
Gas connection G½” -
Customized OEM Solutions
Optimize and simplify your processes with our customized & modular flow solutions.
For a wide range of applications the thermal mass flow meters and mass flow controllers offer you a significant optimization and simplification of your processes.
Measuring & control | |
Ranges Pressure | |
Pressure control | Absolute, differential or gauge pressure Standard measuring ranges from 30 mbar up to 10 bar (graded) Turndown ratio: 1 : 100 |
Back pressure control | Absolute, differential or gauge pressure Standard measuring ranges from 30 mbar up to 10 bar (graded) Dynamic range depending on the application |
Measuring ranges flow | |
Air/Full scale freely selectable | Body size ¼” from 0 … 25 mln/min to 0 … 60 ln/min Body size ½” from 0 … 60 ln/min to 0 … 450 ln/min |
Turndown ratio & accuracy | |
Flow | |
Standard | Accuracy: ± 1.0% of full scale* Turndown ratio: 1 : 50 |
Hi-Performance (up to 150 ln/min) | Accuracy: ± 0.3% of full scale + ± 0.5% of reading* Turndown ratio: 1 : 100 |
Pressure controller with external transmitter, special measuring ranges (e.g. 0-20 Pa) & customer-specific solutions on request | |
Performance Data | |
Media (real gas calibration) | Air*, O2*, N2, He, Ar, CO2, H2, CH4, C3H8 (other gases and gas mixtures on request) *O2 & N2 are calibrated with air |
Response time Flow Measurement | ± 80ms depending on device configuration & according to SEMI standard E17-1011, 5-100% of range under optimized conditions |
Response time Pressure Measurement | 150ms |
Response time Pressure Control | Depending on the measuring section |
Repeatability | ± 0.2% of full scale (according to SEMI standard E56-0309) |
Longterm stability | < 1% of measured value / year |
Power supply | 24 Vdc (18 – 30 Vdc), 15 Vdc on request |
Current consumption Standard | Meter: max. 100 mA; Controller: max. 250 mA (with valve type 8 max. 490mA) |
Current consumption Profinet RT / EtherCAT | Meter: max. 125 mA; Controller: max. 340 mA (with valve type 8 max. 560mA) |
Temperature (environment/gas) | 0 – 50°C |
Body Materials | Anodized aluminium, optional stainless steel electropolished 1.4305 or 1.4404 |
Seals | FKM, EPDM, optional FFKM |
Pressure | Vacuum up to 10 bar g |
Pressure sensitivity | <0.2% / bar of reading (typical N2) |
Temperature sensitivity | 0.025% FS measuring range type / °C |
Warm-up time | <1 sec. for full accuracy |
Integration | |
Output signals digital | RS-485; Modbus RTU (Slave); Lab View-VIs available Option: ProfiBus DP-V0, DP-V1 / Profinet RT / EtherCAT |
Output signals analog | 0..20 mA, 4..20 mA, 0..5 V, 1..5 V, 0..10 V, 2..10 V |
Analog setpoints | Realizable with AD-converter (on request) |
Process connection | G¼” (BSPP* female) up to 60 ln/min, G½” (BSPP* female) up to 450 ln/min *British Standard Pipe Parallel |
Inlet section | None required |
Electrical connection | Sub D plug, 9 pole |
Mounting orientation | Any position (consult manufacturer above 5 bar or vertical mounting) |
Safety | |
Test pressure | 16 bara |
Leak rate | < 1 x 10-6 mbar l/s He |
Ingress protection class | IP50 or IP67 |
EMC | EN 61326-1 |